As a temporary solution, you should delete
any unused ports from the Project. Only PLCs currently connected to
the Citect system should have a port defined and used in the I/O
Device form.
Another approach is to use a local IP address, such as 127.0.0.1
port XXX, in the ports form for the devices that are not connected
for the test setup. This will mean that the device will not come
on-line but the other devices will.
Citect has confirmed this to be a problem in CitectHMI/SCADA
versions 5.xx to 6.10. We are researching this problem and will
post new information here as it becomes available.
==UPDATE==
This issue was initially resolved in v6.10
However, the same issue occurred in v7.1 when run on a Vista
Operating System. In this case, the resolution is to set:
[IOServer]
AlwaysCallStopChannel = 1
AlwaysCallStopUnit = 1
These are 0 by default in version 7.10. In version 7.20, these will
be turned on by default.
|